|
Field
Calibration Procedure
This instrument
is battery powered; therefore, it is highly recommended that a fresh
set of 9 volt alkaline batteries be used to operate the unit during
calibration. The calibration of this instrument will remain accurate
throughout the aging cycle of the batteries due to an onboard voltage
regulator circuit. Additionally, the display will turn off and the
"low battery" indictor will light before the accuracy is affected
by a failing battery. Nevertheless, fresh batteries are recommended
during calibration. It is not necessary to calibrate the instrument
each time batteries are replaced.
Be aware that
when field calibrating the 48A, you are also compensating for variations
in the pressure and temperature sensors. The Sensor Package which
is attached to the main body of the vessel is NOT adjustable, and
no attempt should be made to open the package or alter its function.
All calibration adjustments are inside the Model 48A Volume Computer.
Once the calibration has been performed, it is imperative that the
48A Computer and Sensor Package Vessel remain a matched pair Serial
numbers can be found on the back of the Volume Computer and directly
below the transducer cover (12059). The units are sold as matched
pairs from the factory. If a Volume Computer or vessel is mis-matched,
then the calibration procedure will have to be repeated.
Preparation:
1. Remove the
four rubber feet on the back of the computer, exposing the four
screws which hold the unit together.
2. Remove the
four screws using a No. 1 Phillips screwdriver.
3. Unplug the
top half from the bottom half of the unit by pulling firmly and
evenly on both halves.
4. Connect the
two printed circuit boards together using a 24-circuit boards together
using a 24-circuit male to female jumper (12015).
5. Position
the assembly so the display is visible and the trimpots on the Processor
Board are accessible. It is not necessary (or recommended) to remove
either printed circuit board from its mounting.
6. Connect
the Vessel Sensor Package assembly to the Volume Computer using
the cable supplied.
Adjustment
to the Internal Voltage Reference:
Note: This adjustment
requires a 3-1/2 digit digital voltmeter. If one is not available,
skip this part of the procedure and go on to Temperature Calibration.
1. Connect the
minus lead of the DVM to ground (see Fig.1).
2. Connect the
plus lead of the DVM to TP-1 at lower end of 10K resistor ( see
Fig. 1.) Figure 1

3. Turn the
Volume Computer on by pressing the PRESS mode switch.
4. Adjust the
REF trimpot for a reading of exactly +1.920 volts.
Temperature
Calibration:
Note: A controlled
temperature chamber equipped with a calibrated laboratory thermometer
or temperature monitoring system is required for this adjustment.
The usable temperature range of the Series 12,000 Volume Computer
is from +26.0 degrees F. To +57.0 degrees F. Therefore, the test
temperatures must be within this range. At all times, the Volume
Computer itself should be kept at room temperature. Only the vessel
assembly (sensor package) need to be located inside the temperature
chamber.
To Calibrate:
1. Place the
vessel in the temperature chamber and run the temperature down to
+28F.
2. Allow the
assembly to stabilize at this temperature for at least 20 minutes.
3. Press the
TEMP mode switch on the Volume Computer and check the reading. If
it is not the same as the reference thermometer, adjust the SPAN
trimpot for the correct reading.
4. Run the temperature
chamber up to +55 F.
5. Allow the
assembly to stabilize for at least 20 minutes.
6. Press the
TEMP.. mode switch on the Volume Computer and check the reading.
It if is not the same as the reference thermometer, adjust the ZERO
trimpot for the correct reading.
7. These two
adjustments are slightly interactive; therefore, it is necessary
to repeat the procedure (steps 1 through 6 above) until no further
correction is required.
Pressure
Calibration:
Note; Pressure
calibration requires a compressed air (or pressurized liquid) source
which is equipped with a highly accurate readout system, accurate
and readable to +.1 lb.
1. Apply 0.5
psi pressure to the sensor vessel (Go in through either valved port.
Use the other for bleeding).
2. Select the
PRESS mode on the Volume Computer.
3. Check the
reading. If it does not match the applied pressure, adjust the NULL
trimpot for the correct reading.
4. Increase
the pressure to 24 psi.
5. Check the
reading. If it does not match the applied pressure, adjust the GAIN
trimpot for the correct reading.
6. Bleed the
pressure down to 0.5 psi and re-check the reading. A few cycles
may be required as the adjustments are slightly interactive.
Re-assemble
the Volume Computer in reverse sequence from that called out in
the preparation section. Exercise care not to disturb the trimpot
settings.
Care of the
Instrument:
A) After the
test has been completed open both inlet and outlet valves (12020-B)
and blow out the beer sample with pressure from the aspirator bulb
(12042).
B) Rinse the
inside of the Volume Meter with water to remove any traces of beer.
This can be accomplished by attaching a hose from a water source
to the gooseneck (12045) and then opening the inlet and outlet valves
to let water flush through the instrument. Do not use hot water
as it will damage the thermometer.
C. Apply a small
amount of "O" ring lubricant to the piston (12005) and work the
piston up and down to lubricate the internal piston "O" ring (12009).
D. Rinse out
any traces of beer from the aspirator bulb (12042) to prevent the
check valve from sticking.
E. If the computer
becomes soiled, clean the case with a soft damp cloth and dry. Do
not submerge the computer in water.
|